The development and validation of a system for multi-site photoplethysmography (PPG) and electrocardiography (ECG) is presented. The system could acquire signals from 8 PPG probes and 10 ECG leads. Each PPG probe was constituted of a light-emitting diode (LED) source at a wavelength of 940 nm and a silicon photomultiplier (SiPM) detector, located in a back-reflection recording configuration. In order to ensure proper optode-to-skin coupling, the probe was equipped with insufflating cuffs. The high number of PPG probes allowed us to simultaneously acquire signals from multiple body locations. The ECG provided a reference for single-pulse PPG evaluation and averaging, allowing the extraction of indices of cardiovascular status with a high signal-to-noise ratio. Firstly, the system was characterized on optical phantoms. Furthermore, in vivo validation was performed by estimating the brachial-ankle pulse wave velocity (baPWV), a metric associated with cardiovascular status. The validation was performed on healthy volunteers to assess the baPWV intra- and extra-operator repeatability and its association with age. Finally, the baPWV, evaluated via the developed instrumentation, was compared to that estimated with a commercial system used in clinical practice (Enverdis Vascular Explorer). The validation demonstrated the system’s reliability and its effectiveness in assessing the cardiovascular status in arterial ageing.
Multi-Site Photoplethysmographic and Electrocardiographic System for Arterial Stiffness and Cardiovascular Status Assessment
Perciavalle, Vincenzo;
2019-01-01
Abstract
The development and validation of a system for multi-site photoplethysmography (PPG) and electrocardiography (ECG) is presented. The system could acquire signals from 8 PPG probes and 10 ECG leads. Each PPG probe was constituted of a light-emitting diode (LED) source at a wavelength of 940 nm and a silicon photomultiplier (SiPM) detector, located in a back-reflection recording configuration. In order to ensure proper optode-to-skin coupling, the probe was equipped with insufflating cuffs. The high number of PPG probes allowed us to simultaneously acquire signals from multiple body locations. The ECG provided a reference for single-pulse PPG evaluation and averaging, allowing the extraction of indices of cardiovascular status with a high signal-to-noise ratio. Firstly, the system was characterized on optical phantoms. Furthermore, in vivo validation was performed by estimating the brachial-ankle pulse wave velocity (baPWV), a metric associated with cardiovascular status. The validation was performed on healthy volunteers to assess the baPWV intra- and extra-operator repeatability and its association with age. Finally, the baPWV, evaluated via the developed instrumentation, was compared to that estimated with a commercial system used in clinical practice (Enverdis Vascular Explorer). The validation demonstrated the system’s reliability and its effectiveness in assessing the cardiovascular status in arterial ageing.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.