BACKGROUND Permanent pacemaker implantation (PPI) remains one of the main drawbacks of transcatheter aortic valve replacement (TAVR), but scarce data exist on PPI after valve-in-valve (ViV) TAVR, particularly with the use of newer-generation transcatheter heart valves (THVs).OBJECTIVES The goal of this study was to determine the incidence, factors associated with, and clinical impact of PPI in a large series of ViV-TAVR procedures.METHODS Data were obtained from the multicenter VIVID Registry and included the main baseline and procedural characteristics, in-hospital and late (median follow-up: 13 months [interquartile range: 3 to 41 months]) outcomes analyzed according to the need of periprocedural PPI. All THVs except CoreValve, Cribier-Edwards, Sapien, and Sapien XT were considered to be new-generation THVs.RESULTS A total of 1,987 patients without prior PPI undergoing ViV-TAVR from 2007 to 2020 were included. Of these, 128 patients (6.4%) had PPI after TAVR, with a significant decrease in the incidence of PPI with the use of new-generation THVs (4.7% vs. 7.4%; p = 0.017), mainly related to a reduced PPI rate with the Evolut R/Pro versus CoreValve (3.7% vs. 9.0%; p = 0.002). There were no significant differences in PPI rates between newer-generation balloon-and self expanding THVs (6.1% vs. 3.9%; p = 0.18). In the multivariable analysis, older age (odds ratio [OR]: 1.05 for each increase of 1 year; 95% confidence interval [CI]: 1.02 to 1.07; p = 0.001), larger THV size (OR: 1.10; 95% CI: 1.01 to 1.20; p = 0.02), and previous right bundle branch block (OR: 2.04; 95% CI: 1.00 to 4.17; p = 0.05) were associated with an increased risk of PPI. There were no differences in 30-day mortality between the PPI (4.7%) and no-PPI (2.7%) groups (p = 0.19), but PPI patients exhibited a trend toward higher mortality risk at follow-up (hazard ratio: 1.39; 95% CI: 1.02 to 1.91; p = 0.04; p = 0.08 after adjusting for age differences between groups).CONCLUSIONS In a contemporary large series of ViV-TAVR patients, the rate of periprocedural PPI was relatively low, and its incidence decreased with the use of new-generation THV systems. PPI following ViV-TAVR was associated with a trend toward increased mortality at follow-up. (J Am Coll Cardiol 2021;77:2263-73) (c) 2021 by the American College of Cardiology Foundation.

Permanent Pacemaker Implantation Following Valve-in-Valve Transcatheter Aortic Valve Replacement: VIVID Registry

Barbanti, Marco;
2021-01-01

Abstract

BACKGROUND Permanent pacemaker implantation (PPI) remains one of the main drawbacks of transcatheter aortic valve replacement (TAVR), but scarce data exist on PPI after valve-in-valve (ViV) TAVR, particularly with the use of newer-generation transcatheter heart valves (THVs).OBJECTIVES The goal of this study was to determine the incidence, factors associated with, and clinical impact of PPI in a large series of ViV-TAVR procedures.METHODS Data were obtained from the multicenter VIVID Registry and included the main baseline and procedural characteristics, in-hospital and late (median follow-up: 13 months [interquartile range: 3 to 41 months]) outcomes analyzed according to the need of periprocedural PPI. All THVs except CoreValve, Cribier-Edwards, Sapien, and Sapien XT were considered to be new-generation THVs.RESULTS A total of 1,987 patients without prior PPI undergoing ViV-TAVR from 2007 to 2020 were included. Of these, 128 patients (6.4%) had PPI after TAVR, with a significant decrease in the incidence of PPI with the use of new-generation THVs (4.7% vs. 7.4%; p = 0.017), mainly related to a reduced PPI rate with the Evolut R/Pro versus CoreValve (3.7% vs. 9.0%; p = 0.002). There were no significant differences in PPI rates between newer-generation balloon-and self expanding THVs (6.1% vs. 3.9%; p = 0.18). In the multivariable analysis, older age (odds ratio [OR]: 1.05 for each increase of 1 year; 95% confidence interval [CI]: 1.02 to 1.07; p = 0.001), larger THV size (OR: 1.10; 95% CI: 1.01 to 1.20; p = 0.02), and previous right bundle branch block (OR: 2.04; 95% CI: 1.00 to 4.17; p = 0.05) were associated with an increased risk of PPI. There were no differences in 30-day mortality between the PPI (4.7%) and no-PPI (2.7%) groups (p = 0.19), but PPI patients exhibited a trend toward higher mortality risk at follow-up (hazard ratio: 1.39; 95% CI: 1.02 to 1.91; p = 0.04; p = 0.08 after adjusting for age differences between groups).CONCLUSIONS In a contemporary large series of ViV-TAVR patients, the rate of periprocedural PPI was relatively low, and its incidence decreased with the use of new-generation THV systems. PPI following ViV-TAVR was associated with a trend toward increased mortality at follow-up. (J Am Coll Cardiol 2021;77:2263-73) (c) 2021 by the American College of Cardiology Foundation.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11387/157591
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 19
  • ???jsp.display-item.citation.isi??? 16
social impact