Prototypes of new passivated implanted planar silicon detectors, obtained for the first time from 6 in. silicon slices, have been tested. The time and energy resolutions have been studied as a function of the type and energy of the detected particles, in order to test the performances of these detectors for time of flight measurements in the Chimera project. Some problems arising from edge effects observed in double-pad detectors have been solved by using a guard ring.

Timing performances and edge effects of detectors worked from six inch silicon slices

LANZALONE, GAETANO
Membro del Collaboration Group
;
1997-01-01

Abstract

Prototypes of new passivated implanted planar silicon detectors, obtained for the first time from 6 in. silicon slices, have been tested. The time and energy resolutions have been studied as a function of the type and energy of the detected particles, in order to test the performances of these detectors for time of flight measurements in the Chimera project. Some problems arising from edge effects observed in double-pad detectors have been solved by using a guard ring.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11387/7306
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 14
  • ???jsp.display-item.citation.isi??? 17
social impact