Mass and atomic-number identification (ID) of reaction products is a fundamental requirement of any nuclear reaction study. An effective particle-ID method is demonstrated, based on pulse shape analysis/discrimination (PSD) applied to large-area, single-element silicon detectors. This technique uses commercial electronic modules and achieves atomic number resolution rivaling that typically obtained with multi-element (ΔE - E) detector telescopes. The method is applied to the CHIMERA detector system without compromising its time-of-flight (TOF) resolution. In-beam tests of the PSD method have been performed with large-area, 300-μm thick CHIMERA silicon detectors, measuring particles from the 19F + 12C reaction at Tandem energies. Performance of a simple PSD set up is discussed, for front and rear particle injection.
Titolo: | Particle Identification via Pulse Shape Analysis for Large-Area Silicon Detector of the CHIMERA Array |
Autori: | |
Data di pubblicazione: | 2005 |
Rivista: | |
Handle: | http://hdl.handle.net/11387/8304 |
Appare nelle tipologie: | 1.1 Articolo in rivista |